@sontakey/ppg-js
v0.3.0
Published
Camera PPG in the browser: heart rate, pulse-rate variability, respiration and signal quality from a fingertip on a phone camera. Zero dependencies, typed, no UI.
Downloads
358
Maintainers
Readme
ppg-js
Heart rate, pulse-rate variability, respiration and an honest signal-quality gate from a fingertip on a phone camera, in the browser. Zero dependencies, typed, no UI.
ppg-js turns the camera (and the flash, where the browser exposes it) into
a photoplethysmography sensor: cover the rear lens with a fingertip and it
reports heart rate, inter-beat intervals, RMSSD/SDNN, breathing rate and a
per-window quality verdict that says why a number is withheld rather than
guessing. Every session is recorded and can be replayed offline through the
exact same engine.
Live demo: https://ppg-js.vercel.app (needs a phone and HTTPS)
The demo app in examples/app is the three-minute HRV Spot Check shown
above: start, measuring
(screen recording from an iPhone), and the report
(frequency domain, full page)
rendered from a real recording in test/fixtures/.
Contents
- Install
- Quick start
- What you get
- How it works
- API
- Signal quality contract
- HRV module
- Other sources
- Browser and device support
- Debug recording and replay
- Accuracy and limitations
- Testing
- Contributing and license
Install
npm install @sontakey/ppg-jsBrowser global (no bundler):
<script src="https://unpkg.com/@sontakey/ppg-js/dist/index.global.js"></script>
<script>
const ppg = new PPG.PPG();
ppg.addEventListener('metrics', (e) => console.log(e.detail.heartRate, e.detail.quality));
document.querySelector('#start').onclick = () => ppg.start();
</script>The global build exposes PPG.PPG, PPG.PPGMonitor, PPG.PpgEngine,
PPG.hrv, PPG.dsp and PPG.sources.
Quick start
import { PPG } from '@sontakey/ppg-js';
const ppg = new PPG();
ppg.addEventListener('state', (e) => {
// NO_FINGER -> SETTLING -> MEASURING
console.log(e.detail.state, e.detail.reason);
});
ppg.addEventListener('metrics', (e) => {
const m = e.detail;
if (!m.quality.good) {
console.log('not ready:', m.quality.reason, m.guidanceMessage);
return;
}
console.log(`${m.heartRate} bpm, RMSSD ${m.rmssd.toFixed(0)} ms (noise floor ±${m.rmssdFloorMs.toFixed(0)} ms)`);
});
ppg.addEventListener('respiration', (e) => console.log(`${e.detail.rateBpm.toFixed(0)} breaths/min`));
ppg.addEventListener('error', (e) => console.error(e.detail.error.code, e.detail.error.guidance));
// Call from a user gesture (tap): camera, wake lock and motion permission all need one.
button.onclick = () => ppg.start();
// later
ppg.stop();After a session:
import { hrv } from '@sontakey/ppg-js';
const beats = ppg.getTachogram({ goodOnly: true, hrvOnly: true }).filter(b => b.valid);
const report = hrv.analyzeHRV(beats, { rmssdFloorMs: ppg.getMetrics().rmssdFloorMs });
console.log(report.timeDomain.rmssd, report.frequencyDomain.lfhf, report.stressIndex.sqrt);What you get
| Field | Meaning |
|---|---|
| heartRate | bpm, median of the last accepted intervals, cross-checked against a spectral estimate; 0 unless quality.good |
| ibi, rmssd, sdnn | most recent interval and variability over the last 60 s of accepted beats; 0 unless quality.good |
| rmssdFloorMs, timingUncertaintyMs | the RMSSD the pipeline's own beat-timing noise would produce on a perfectly regular pulse, estimated per window (accurate to roughly ±50%) |
| respiration | { rateBpm, confidence, basis } from breathing-driven modulation of beat timing, pulse amplitude and baseline (5.5-30 breaths/min). Confidence ≥ 0.6 when at least two estimates agree, 0.5 for a pair with one weak source, 0.3 for a single clear source (provisional), 0.4 while a recent firm rate is held. First estimate about 45 s after MEASURING |
| quality | { good, reason, code }: the gate verdict, why, and a stable code (see the contract below). Its inputs sit beside it: acDcRatio, artifactRatio, ibiFftDisagree, clippedFraction, motion, templateSqi |
| templateSqi | median correlation of each pulse with the window's mean pulse shape (0-1) |
| acDcRatio, perfusionIndex | pulsatile amplitude over DC of the selected channel (bandpassed per-beat, not raw range) |
| heartRateFFT, heartRateIBI, heartRateSource, harmonicCorrected | the two independent estimates and which one won |
| fingerState, settleRemainingSec, guidanceMessage | state machine and coaching copy |
| sampleRate, selectedChannel, redDc, greenDc, clippedFraction, motion | capture diagnostics |
Every beat event carries { time, ibiMs, valid, good, lowSnr, sqi, reason, heartRate }: valid is
the interval-level artifact decision, good is whether the window it came
from passed the quality gate, and lowSnr marks an interval next to a
recovered weak beat (counted for heart rate, excluded from variability).
How it works
camera frame (requestVideoFrameCallback, captureTime when available)
-> centre ROI, downscaled, channel means + clipped-pixel fraction
-> PpgEngine.push({ t, r, g, b, clipped, motion })
-> finger state machine every sample (relative presence, drift)
-> every 5 s of signal time:
interpolate the last 8 s onto one absolute 60 Hz grid
zero-phase Butterworth bandpass (4th-order HP 0.6 Hz, 2nd-order LP 4.6 Hz)
channel selection by pulsatile amplitude
spectral HR (Hann, zero-padded, sub-harmonic guard) -> refractory prior
peaks (adaptive threshold + 0.3 s vertex fit) -> absolute beat times
missed-beat recovery (weak pulse inside a 2x gap, flagged lowSnr)
template correlation per beat -> interval validation over the continuous stream
(range, missed-beat multiples, jump vs the recent median; four consistent
rejects re-seed the median so a noisy start or a real rate change is followed)
RMSSD / SDNN over 60 s, respiration from beat modulation
quality gate -> good / reason / codeThe same PpgEngine runs the live camera path and tools/replay.js, so a
recorded session replays to identical windows (test/live-parity.test.js
asserts it).
Finger state machine
NO_FINGER --finger placed--> SETTLING --6 s, low drift, pulse amplitude ok--> MEASURING
^ | |
| finger lifted large drift
+----------------------------+----------------------------------------------+Exposure, white balance and focus are locked when MEASURING is first reached (the auto algorithms have converged on the finger by then) and released when the finger lifts.
API
new PPG(options?)
Typed-event facade, always headless. Events: ready, state, beat,
metrics, quality, waveform (every frame), respiration, error.
Methods: start(abortSignal?), stop(), destroy(), getMetrics(),
getTachogram({ goodOnly, hrvOnly }), getSessionSummary(), exportDebugLog(),
downloadDebugLog(prefix?), getConfig(); getters state, capabilities,
engine.
new PPGMonitor(container, options?)
The callback-style class the facade wraps (container is accepted for
compatibility and ignored; the core renders no UI). Callbacks: onReady,
onState, onQualityUpdate (per window), onSignalUpdate (per frame),
onFrame, onError. Same methods as PPG plus getDebugLog(),
copyDebugLogToClipboard(), getEngine(), and the video element as a
property for previews.
Options (all optional)
{
signal: { // PpgEngine options (the common ones; see EngineOptions for all)
windowSec: 5, // analysis window, seconds of signal time
hopSec: 5, // interval between windows
contextSec: 3, // extra history filtered with each window
gridHz: 60, // uniform analysis grid
hrSlewPerWindow: 8, // max displayed HR change between consecutive good windows
recoverMissedBeats: true, // look for a weak beat inside a 2x gap (flagged lowSnr)
cardiacBandLow: 0.75, // Hz, heart-rate search band (45 bpm)
cardiacBandHigh: 4.0, // Hz (240 bpm)
quality: { minAcDc: 0.002, maxArtifactRatio: 0.2, minIbiCount60s: 8,
maxClippedFraction: 0.05, maxMotion: 1.5, minTemplateSqi: 0.6 },
fingerState: { settleSec: 6, driftEnterFraction: 0.03, driftExitFraction: 0.06,
presence: { minRed: 60, minRedShare: 0.5 } },
respiration: true, templateSqi: true
},
camera: {
width: { ideal: 640 }, height: { ideal: 480 }, frameRate: { ideal: 60 },
facingMode: 'environment',
torch: true, // request the flash when the device advertises it
lockExposure: 'measuring', // 'measuring' | 'start' | 'never'
zoom: null // opt-in digital zoom (some Androids switch lens above 1)
},
roi: { widthFraction: 0.3, heightFraction: 0.3 },
wakeLock: true, // keep the screen on while measuring
motion: false, // DeviceMotion gate (asks permission on iOS)
debug: { persistLastSession: false, includeUserAgent: false, sampleCap: 36000 },
video: null // supply your own <video> element
}Errors
start() rejects with a PPGError whose code is one of
insecure_context, unsupported, permission_denied, no_camera,
camera_busy, overconstrained, aborted, unknown, and whose guidance
is a sentence you can show to the user.
PpgEngine (@sontakey/ppg-js/engine)
Feed it your own samples: engine.push({ t, r, g, b, clipped?, motion? })
returns { state, stateChanged, stateReason, present, waveform, window } (window is null except every hopSec). Use it for other
capture paths (a WebView bridge, a wearable's raw PPG, a file).
DSP (@sontakey/ppg-js/dsp)
Pure functions: designBandpass, filtfilt, filtfiltPadded,
resampleToGrid, computeFFT, calculateSNRFromPSD, detectPeaksDetailed,
computeIBIs, crossCheckHeartRate, templateCorrelation,
estimateRespiration, evaluateQuality, FingerStateMachine.
Signal quality contract
quality.good is true only when, in this order, all of these hold; reason
and code name the first that fails:
fingerState === 'MEASURING'(no_finger,settling)- clipped-pixel fraction ≤ 5% (
saturated) - device motion ≤ threshold, when a motion source is attached (
motion) - pulsatile AC/DC ≥ 0.2% (
weak_pulse) - ≤ 20% of candidate intervals rejected in the last 60 s (
irregular); intervals from before the beat reference settled after a rough start count as settling, not rejects - ≥ 8 accepted intervals in the last 60 s (
collecting) - median template correlation ≥ 0.6 (
morphology) - interval-based and spectral heart rate agree within 25%
(
double_count,missed_beats,fft_disagree)
heartRate, rmssd and sdnn are 0 whenever good is false;
heartRateRaw is always populated.
Signal quality indices
Every window carries metrics.sqi, the standard indices from the PPG
quality literature plus one documented composite, so you can build your own
acceptance rule or compare with other toolkits:
| Field | Definition |
|---|---|
| score | 0-1 composite: geometric mean of bounded sub-scores for template correlation, artifact ratio, pulse amplitude, SNR, clipping, motion and detector agreement (components lists each) |
| skewness, kurtosis | of the bandpassed pulse (Elgendi 2016: skewness is the most informative single index; a clean pulse is positively skewed) |
| perfusion | pulsatile amplitude / DC, percent |
| relativePower, snrDb | cardiac-band power over total power, linear and in dB |
| zeroCrossingRate | zero crossings per second of the bandpassed pulse (about 2 per beat when clean) |
| templateCorrelation | median per-beat correlation with the window's mean beat (Orphanidou 2015) |
| detectorAgreement | 1 − |HR from intervals − HR from spectrum| / HR from spectrum |
| artifactRatio, clippedFraction, motion | the gate's inputs |
Each beat event and tachogram point also carries sqi, that beat's
template correlation (the lower of its two peaks), so beats can be weighted
or filtered individually. The composite is for ranking and display; the
accept/reject decision remains quality.good.
HRV module
@sontakey/ppg-js/hrv works on any list of intervals (numbers in ms, or
{ ibiMs, t } with beat times so gaps are handled correctly):
timeDomain: mean RR/HR, SDNN, RMSSD, lnRMSSD, NN50/pNN50, min/max HR, triangular index, least-squares TINN, RR rangefrequencyDomain: Welch PSD on a 4 Hz linear resample using real beat times, LF/HF (VLF only for ≥ 5 min), normalised units, HF-peak respiration, coherence (0.04-0.26 Hz peak share); refuses recordings with > 20% gaps. PassrespirationRateBpm(the engine's fusion estimate) andrespirationInLftells you when breathing under 9/min has moved RSA into the LF band, so LF/HF is breathing-driven rather than sympatheticnonlinear: Poincaré SD1/SD2, sample entropy, DFA α1stressIndex: Baevsky SI and its square root (the form Kubios reports)ansIndices(experimental): PNS/SNS z-scores against cited referencesultraShortRmssd(last 60 s),lnRmssdBaseline(7-day rolling mean, CV, smallest-worthwhile-change band)analyzeHRV: all of the above
What a camera measures is pulse-rate variability. At rest it tracks heart-rate variability closely; under posture or temperature change the two diverge. Treat the numbers accordingly.
Other sources
@sontakey/ppg-js/sources has BleHeartRateSource (Web Bluetooth Heart
Rate Service: RR intervals from a Polar/Garmin chest strap, Chrome on
Android and desktop) and ArraySource (recorded samples through the
engine). Recording a strap alongside the camera is the cheapest way to
validate this library on your own device.
Browser and device support
| Platform | Camera | Flash (torch) | Notes |
|---|---|---|---|
| iOS Safari 17+ and every iOS browser (WebKit) | yes | yes | exposureMode is not exposed on iPhone rear cameras; white balance is. The engine's late lock handles both. |
| Android Chrome | yes | yes | Torch, exposure/white-balance/focus locks, zoom, Web Bluetooth. Lens selection by label is best-effort. |
| Desktop Chrome/Edge/Firefox | yes | no | Works with a desk lamp for development; not a measurement setup. |
| In-app browsers (Instagram, Facebook, some WebViews) | often no | no | start() rejects with unsupported. |
Requirements: HTTPS (or localhost), getUserMedia. requestVideoFrameCallback
is used when available; requestAnimationFrame otherwise.
Tested against six real recordings from one iPhone (iOS 26, Safari and
Chrome, 60 fps, in test/fixtures/: clean, muted, slow breathing, periodic
30 fps dips, a noisy onset) and against a simulator at 24/30/60 fps with
dropped frames, saturation, motion, respiratory modulation and dominant
dicrotic waves, and benchmarked against HeartPy and vital_sqi
(docs/audit/BENCHMARK-2026-09.md). Android recordings are
the most valuable thing you can contribute right now; see
CONTRIBUTING.md.
Debug recording and replay
Every session records per-frame channel means, timestamps, camera
capabilities and settings, applied constraints, dropped-frame counts, every
peak and interval decision, and per-window quality records.
getDebugLog() returns it; downloadDebugLog() saves it. Nothing is written
to localStorage unless debug.persistLastSession is set.
npm run replay -- path/to/ppg-debug.jsonreplays the recording through the same engine the browser used and prints state transitions, per-window quality, the HR timeline with the RMSSD floor, every interval decision, respiration, and a comparison with the peaks the phone recorded live.
Accuracy and limitations
Measured on the simulator (what the tests enforce):
| Case | Result | |---|---| | Heart rate, 60-90 bpm, 24/30/60 fps | within 1 bpm | | RMSSD floor on a zero-variability pulse, fractional ROI means | 9-13 ms (bound 15 ms) | | Same with whole-count quantised means | 11-16 ms (bound 22 ms) | | 10% dropped frames | raises RMSSD by < 2 ms (bound 5 ms) | | Respiration at 6 breaths/min | 6.0 ± 1 |
And on the real recordings (what test/fixtures.test.js enforces):
| Case | Result | |---|---| | Live vs replay, 3 min session | 212 of 212 beats reproduced | | Time to first heart rate, clean start | 10 s after MEASURING (20 s after finger placement: 6 s settle, then 8 accepted beats) | | Time to first heart rate, noisy onset (junk intervals seeded the reference) | 15 s after MEASURING (25 s after placement); before 0.3.0 the session never recovered | | Periodic 30 fps dips (every other frame dropped for 20 s) | HR unchanged, RMSSD floor 7 -> 14 ms during the dip | | Breathing at 6.7/min | first estimate 40 s after MEASURING, then available in every window but one |
Not validated: no ECG or chest-strap comparison across a population, no
claim across skin tones or ambient light beyond the recordings in
test/fixtures/. This is not a medical device. It is for research and
personal curiosity, not diagnosis, treatment, or any decision that needs a
validated instrument.
Testing
npm test # strict typecheck + all unit suites
npm run test:e2e # real Chromium + fake camera fed from a fixture (after npm run build)Contributing and license
See CONTRIBUTING.md. MIT, see LICENSE.
The audit that drove the 0.3 rebuild, with its measurements and references,
is in docs/audit/AUDIT-2026-09.md.